Accuracy of Pulse Oximeters Intended for Hypoxemic Pediatric Patients - Case study

SHARE
May. 19, 2022- By: Ramamoorthy C;Shiboski SC
Courtesy ofMasimo Corp

OBJECTIVES: Prior studies have shown inaccuracies in pulse oximetry readings at saturations less than 85%; however, no large studies have evaluated new sensors marketed for these low saturations. This study`s purpose was to evaluate two sensors with claims of improved accuracy in children with saturations less than 85%.

DESIGN: Prospective observational study.

SETTING: Single institution; cardiac catheterization laboratory, and operating room.

PATIENTS: Fifty patients weighing 3-20 kg with baseline saturations less than 90% undergoing surgical or catheterization procedure.

MEASUREMENTS AND MAIN RESULTS: Data collected included demographics, diagnosis, continuous saturations from three different pulse oximeters (Masimo LNCS [Masimo, Irvine, CA], Masimo Blue [Masimo], and Nellcor Max-I [Medtronic, Dublin, Ireland]) and up to four blood samples for co-oximetry as the gold-standard arterial oxygen saturation. Analysis included scatter plots, smoothed regression estimates of mean continuous saturation levels plotted against corresponding arterial oxygen saturation values, and Bland-Altman plots. Bland-Altman analysis indicated increasing levels of bias and variability for decreasing arterial oxygen saturation levels for all three sensors, with a statistically significant increase in mean difference observed for decreasing arterial oxygen saturation level. The Masimo Blue sensor had the lowest mean difference, SD and Bland-Altman limits in patients with saturations less than or equal to 85%. At saturation range of less than or equal to 85% and greater than 75%, 14% of the samples obtained from Masimo Blue, 24% of the readings from the Nellcor, and 31% from the Masimo Standard sensors were greater than or equal to 5% points difference. All three sensors had a further increase in these differences for arterial oxygen saturation values less than 75%.

CONCLUSIONS: The Masimo Blue sensor has improved accuracy at saturations 75-85% versus the Nellcor and Masimo Standard sensors. The accuracy of peripheral capillary oxygen saturation of the Masimo Blue sensor was within 5% points of the arterial oxygen saturation the majority of the time. Currently, at saturations less than or equal to 85%, pulse oximetry alone should not be relied on in making clinical decisions.