Checking a SenTec TC Sensor: Check the condition of the sensor membrane and the integrity of the sensor before and after each use and after changing the membrane!
According to the procedure described point by point below, a Multi-Site Attachment Ring (MAR) or Multi-Site Attachment Ring Easy (MARe) is first attached to the measurement site, one small drop of contact liquid is then applied to the center of the sensor and finally the sensor is snapped into the ring. Alternatively, you may also click the sensor into the ring first, remove the protective liner of the adhesive tape, and then apply one small drop of contact liquid on the center of the sensor. In this case...
According to the procedure described point by point below, a Multi-Site Attachment Ring (MAR) or Multi-Site Attachment Ring Easy (MARe) is first attached to the measurement site, one small drop of contact liquid is then applied to the center of the sensor and finally the sensor is snapped into the ring. Alternatively, you may also click the sensor into the ring first, remove the protective liner of the adhesive tape, and then apply one small drop of contact liquid on the center of the sensor. In this case...
According to the procedure described point by point below, the Ear Clip is first attached to the earlobe, then one small drop of contact liquid is applied to the sensor surface, and, finally, the sensor is snapped into the Ear Clip attached to the earlobe. Altematively, you may click the sensor into the clip`s retainer ring first, apply one small drop of contact liquid to the center of the sensor, pull off both liners protecting the clip`s adhesive tapes, and then apply the sensor/clip assembly to the earlobe....
The membrane of a SenTec TC Sensor must be changed if the `Membrane Change Interval` has elapsed. In this case, the SDM displays the message `Change sensor membrane`, triggers a low priority alarm, marks PCOa/P02 as invalid and activates the menu `Membrane Change` - provided the sensor is in the Docking Station. Additionally, the sensor membrane must be changed if it is damaged, not properly seated, or if there is trapped air or dry electrolyte under the membrane.