The medical gas console outlet shall be Ohio Medical*, Healthcair®Series. The outlets shall be UL Listed, NFPA 99 compliant, cleaned for medical gas service and be pressure tested. Each oulet shall have less than 3 psi (21 kPa) pressure drop through the oulet @ 120 l/min, and 50 psi (345 kPa) inlet pressure. For outlets providing positive ...