Diamond Scientific - Authorized Distributor
G200-00N N2O Analyzer For Medical Applications - Technical Specification
G100
TECHNICAL SPECIFICATION
POWER SUPPLY
Battery type Li Ion
Battery life 10 Hours (8 hours with pump)
Battery lifetime >300 Cycles
Battery charger 5v DC external power supply and internal charging circuit
Charge time 3 Hours
Alternative power USB connector
DC power supply
GAS RANGES
Gases Measured CO2 By custom dual wavelength infra-red
cell with reference channel
O2 (Optional) By internal electrochemical cell
Oxygen cell lifetime Approximately 3 years in air
Range CO2 0-20%
O2 0-100%
Measurement
Accuracy
Gas
CO2 Accuracy: ±( 1% of range+ 2% of reading) at
reference conditions1
Temperature dependence: ±0.2% reading /oC
(typical at 5% CO2)
Pressure dependence: ±0.02% of reading/hPa
(typical at 5% CO2)
O2 ±1.0% Full Scale @ constant temperature and
pressure
±2.0% Full Scale over operating temperature range
Response time, T90 CO2 =20 seconds
O2 =60 seconds
1 Conditions during factory calibration, typically 20oC, 1000mBar
FACILITIES
Temperature (Optional) X2 using optional probes, range 0°C to +50.0°C
Temperature accuracy,
typical
±0.2°C from 32 to 44oC, ±0.5°C over the rest of the range
Barometric pressure 800-1200 mbar
RH measurement
(Optional)
RH Probe 0-100% RH non condensing
RH accuracy ±1.5%RH across the range
Visual and audible alarm User selectable CO2 and O2 alarm levels
Communications USB type B mini-connector, HID device class
Data Storage 1000 reading sets + 270 events
Page 1 of 2 G100 Technical Specification TS1 Iss.03 - 05/05/2010
Page 2 of 2 G100 Technical Specification TS1 Iss.03 - 05/05/2010
G100
TECHNICAL SPECIFICATION
PUMP
Flow 100cc/min typically
ENVIRONMENTAL CONDITIONS
Operating temperature 5°C – 40°C
Relative humidity 0 - 95% non condensing (RH Probe 0-100% non condensing)
Barometric pressure ±500mbar from calibration pressure
IP rating IP40
PHYSICAL
Weight 495 grams
Size L 165mm, W100mm, D55mm
Case Material ABS / Polypropylene with Silicone Rubber Inserts
Keys 17 Resin capped Silicone rubber keys
Display Liquid crystal display, 128 x 64 pixel
With RGB LED back-light
Gas sample filters Built-in gas dryer tube to remove moisture
User replaceable PTFE water trap filter
" Note: Due to Geotech’s continuous programme of improvement, this specification is