All area and main line isolation valves shall be Ohio Medical® series. Valves shall be either locking or non-locking. The valves shall be oxygen cleaned, full port, three piece ball type with flanges to allow easy service and installation. The valve shall have a bronze body, and a blow-out proof stem. The valve ball shall be bronze chrome plated, and the seats and packing shall be Teflon,S)(PTFE). The valve shall have a pressure rating of 600 psig, and be hydrostaticlly tested. Type `K` copper pipe extensions that are cleaned for medical gas service shall be fitted into each side of the valve. Two guage ports shall be drilled into pipe extensions for the purpose of inserting a guage if required. The valve assembly shall be plugged or capped and sealed in a plastic bag to prevent contamination.