Wall/Ceiling Outlets Brochure
The medical gas wall outlet shall be Ohio Medicaf Series. The outlet shall be UL* Listed, NFPA994* compliant, cleaned for medical gas service, and be pressure tested. Each oulet shall have less than 3 psi (21 kPa) pressure drop through the oulet @ 120 L/min, and 50 psi (345 kPa) inlet pressure. For outlets providing positive pressure gas, the ...