1. products
  2. airborne respiratory pathogen analysis system including sars cov 2

Sentinel -Model 720 -Airborne Respiratory Pathogen Analysis System - Including SARS-CoV-2

SHARE

Indoor air sampling with integrated on-site PCR analysis for airborne coronavirus.  A fast and cost-effective COVID-19 workplace safety solution for: Schools, Factories and food processing plants, Restaurants and entertainment venues, Prisons and detention centers, Retail environments

Most popular related searches

SARS-CoV-2 is airborne, and transmitted asymptomatically.   Frequent testing of indoor air for airborne coronavirus will help make your workplace safer.  Environmental surveillance is complementary and additive to random or voluntary testing for COVID-19 in the workforce.

Collect Samples

Collect air samples from each group of employees (those working indoors and together for extended periods of time)

Test Samples

Test air samples with the BioFire® FilmArray® 2.0

Positive PCR Test

If the PCR test is positive, follow established health department guidance

Negative PCR Test

No action needed

BioFlyte® Aerosol Sampling Technology

Highest performing handheld air sampler on the market.

BioFlyte`s BioCapture z720 is the highest performing handheld air sampler on the market, providing a high-flow rate of 200 liters/minute, with collection onto a filter with predictable sample extraction. Seen in the chart below, the concentration factor is the true figure of merit.

BioFire FilmArray 2.0

A sensitive and accurate PCR system, providing detection of SARS-CoV-2 in less than one hour.

The BioFire system allows rapid, on-site direct measurement of airborne coronavirus in the workplace. Results are available in less than one hour. This enables rapid isolation of work groups for further testing following established health department guidance.

How much airborne virus can be collected?

With a 10-minute sampling time, and a flow rate of 200 liters per minute, the RNA concentration in the collected sample will be:

CL =  Ca * F * T  * ηc / VL        
CL = 1 copy / l * 200 l/min * 10 min * 90%  / 5 mL 
CL = 360 RNA copies/mL

 

BioCapture z720 can concentrate 1 copy per liter of air into 360 copies per ml of sample fluid.
Is it sufficient to detect with PCR?

Where,


CL =  RNA concentration in ambient air (RNA copies /L) 
Ca =  RNA concentration in collection liquid (RNA copies /mL) 
F    =  flow rate (L/min)
T    =  sampling time (min)
ηc  =  overall efficiency of collection
VL  =  volume of collected liquid sample

BioFire FilmArray 2.0 can detect 300 copies per ml.