BEBIG WOmedModel T-105 -Superficial X-Ray Therapy (SXRT) System

SHARE

The WOmed T-105 is a superficial X-ray therapy (SXRT) system designed for the treatment of various dermatological conditions. It is particularly effective for non-malignant dermatological diseases such as psoriasis, itching dermatoses, chronic eczema, and hemangioma, as well as for hypertrophic processes like keloid scars. The system is also utilized for curative irradiation treatments of basal cell carcinoma, spinocellular carcinoma, cutaneous T-cell lymphoma, and other conditions. The WOmed T-105 operates with high tension ranging from 10 to 100 kV, a maximum tube current of 12 mA, and a peak output power of 1,200 W. The system features a two-circuit water-water cooling technology and supports four physical filters for enhanced precision. A range of 14 applicators is available, designed to deliver targeted treatments with varying field sizes and focal-skin distances. The unit is built with a versatile, mobile trolley setup for ease of use and adaptability to clinical settings. The WOmed T-105 is engineered in Germany and distributed through BEBIG Medical, available in select markets.

Most popular related searches

Our superficial X-ray therapy system WOmed T-105 is optimized for superficial dermatological diseases

  • Dermatological non-malignant diseases (Psoriasis, itching dermatoses, chronic eczema, hemangioma)
  • Hypertrophic processes (Keloid Scars)
  • Curative irradiation treatment (Basalioma, Spinocellular carcinoma, Cutaneous T-cell lymphoma (Mycosis fungoides), Cutaneous B-cell lymphoma, Merkel cell carcinoma, Lentigo maligna, Kaposi's Sarcoma)
  • High Tension : 10…100 kV
  • Max. Tube Current : 12 mA
  • Max. Output Power : 1.200 W
  • Cooler-Technology : Two-circuit (water-water)
  • Filter-Changer : Manually, 4 physical filters supported
  • Numbers of selectable clinical energies [kV / mA / Filtration] : 20
  • Standard Energiesetup**[kV / mA / Filtration / 1st HVL] :
    • 10kV / 5mA / 1.0mm Be / 0.10mm Al
    • 30kV / 10mA / 0.5mm Al / 0.36mm Al
    • 50kV / 10mA / 1.0mm Al / 1.00mm Al
    • 100kV / 12mA / 0.5mm Cu / 0.45mm Cu
  • Numbers of Applicators : 14
  • Standard Applicator Set** [field size/ Focal-Skin-Distance in cm] :
    • d1/15, open, max. 50 kV
    • d2/15, open, max. 50 kV
    • d3/15, open, max. 50 kV
    • d4/15, closed, max. 100 kV
    • d5/15, closed, max. 100 kV
    • d7/25, closed, max. 100 kV
    • d10/30, closed, max. 100 kV
  • Stand Option : Mobile trolley with X-ray stand arm connected to Wall box (3m range), manual movements, manual locks
  • Mains : 1~PEN 230 V AC, 16 AT
  • Country of Origin : Germany