Bionime Rightest -Model GM260 -Blood Glucose Meter

SHARE

The GM260 delivers accurate measurements of your blood glucose in just 5 seconds. A large display and auto-coding are featured in this simple and clear design, while Validus test strips ensure an easy, hygienic test experience.

Most popular related searches

Fast Results
Displayed in 5 seconds.

Large Display
Easy to view results.

Auto-Coding
Eliminates the need for manual coding and delivers maximum accuracy.

Tiny Blood Sample
Requires only 0.75µL.

Side-Insert Design
Easy to handle and prevents blood contact for excellent hygiene.

Proprietary Strips
Enhanced size for improved grip, ensuring accuracy and precision.

  • Test strip ; GS260
  • Coding ; Auto-coding
  • Enzyme / Measurement technology ; GOD / Electrochemical sensor
  • Blood sample source ; Capillary whole blood
  • Minimum sample volume ; 0.75 µL
  • HCT range ; 30 - 57%
  • Measuring range ; 0.6 - 33.3 mmol/L (10 - 600 mg/dL)
  • Meter dimensions (L x W x H) ; 98.0 x 45.0 x 16.5 mm (3.85 x 1.77 x 0.65 inch)
  • Display area (L x W) ; 46.0 x 33.5 mm (1.81 x 1.31 inch)
  • Weight (with battery) ; 55.0 ± 5 g
  • Power supply ; 1 × CR2032 battery
  • Memory capacity ; 500 test results with date and time
  • Operating temperature ; 10 - 40°C (50 - 104°F)
  • Operating relative humidity ; 10 - 90%
  • Meter storage conditions ; -10 - 60°C (14 - 140°F)
  • Test strip storage conditions ; At 4 - 30°C (39 - 86°F),and below 90% relative humidity
  • Accuracy ; For glucose concentrations < 5.5 mmol/L (100 mg/dL),
  • 95% of measured glucose values within ± 0.83 mmol/L (15 mg/dL) of reference measurements.
  • For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL),
  • 95% of measured glucose values within ± 15% of reference measurements.
  • Precision ; For glucose concentrations < 5.5 mmol/L (100 mg/dL), SD < 0.28 mmol/L (5 mg/dL)
  • For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), CV < 5%