DiconexModel InCCA Counter 18sc - Clinical Chemistry Autoanalyzer

SHARE

The Clinical Chemistry Autoanalyzer InCCA Counter 18sc is a measuring instrument and characterization of blood cells for in vitro diagnostic use, meassure 18 of the most important blood parameters. The instrument is fully automatic. Autosampler included.

Most popular related searches

PARAMETERS

  • WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, PCT, PDW
  • LYM %, LYM #, MID %, MID #, GRAN % , GRAN #
  • WBC, RBC, PLT histograms

SAMPLE VOLUME

  • 25 µl

MEASUREMENT CHAMBERS

  • 3 (RBC, PLT), (WBC) and Mix chamber

APERTURE DIAMETER

  • 70 µm/ RBC-PLT and 100 µm/WBC

THROUGHPUT

  • 50 Test/hour with Autosampler and 60 Test/hour in manual mode

ACCURACY

  • WBC: 3%, RBC: 2%, HCT: 3%, MCV: 2%, PLT: 4%

VOLUME SENSORS TO:

  • Reagents and waste

MULTILANGUAGE

  • English, Spanish, Portuguese (others available)

CONTROL DATA MANAGEMENT

  • Q.C. graphics: incl. mean, min/max range, SD and CV for all measured and calculated parameters, Levey-Jennings plot, separate QC database
  • Capacity to store 150.000 samples with histograms

DIFFERENT SPECIES SUPPORTED

  • Human
  • Veterinarian (dog, cat, horse, up to 6)

PRINCIPLES

  • Impedance method for WBC, RBC and PLT •Colorimetric method for hemoglobin test

SAMPLING METHODS

  • Autosampler with Cap Piercer for 48 tubes
  • Open or closed tube sample

MAINTENANCE PROGRAM

  • Clog prevention: High-voltage pulses on both apertures (each analysis cycle)
  • Chemical cleaning of both apertures (Cleaner reagent)
  • Positive and negative solution pressure flush mode
  • Internal and external washing probe system

INSTALATION

  • Power requirements: 110 - 220 V/50 - 60 Hz. less than 50 VA

WEIGHT

  • 20 kg

DIMENSIONS

  • Width: 350mm; Height: 470mm; Deep: 380mm

PRINTERS

  • Built in thermal printer

OPERATING ENVIRONMENT

  • Temperature 15 ºC – 30 ºC.