Fluke Biomedical - Model DPM4 - Vacuum Pressure Tester
From Biomedical Test Equipment - Digital Pressure Meters
DPM4 Parameter Tester is a highly accurate meter for testing a wide range of medical devices. Key features include its lightweight, compact size and battery operation. The DPM4 is ideal for the testing done as part of preventive maintenance or repair processes whenever measurement of pressure, flow, or relative humidity is required.
-
Most popular related searches
Key Features
- Palm Size
- High accuracy
- Differential pressure, vacuum, and temperature measurements
- Multiple user-selectable units of measurements
- Simultaneous display of multiple parameter measurements
- Leak detection/leak-rate calculation
- RS232 for computer control
- Pressure measurements from -700 mmHg to +5000 mmHg range (applies to the 1G and 2G models only)
- Barometric-pressure, flow, and humidity measurements (applies to the 2G and 2H models only)
Overview
The DPM4 is available in four models: 1H, 1G, 2H, and 2G. The DPM4-1G is the replacement for the popular, but now obsolete DPM3. Basic parameters cover pressure, vacuum, and temperature measurements (Models 1H and 1G). Gas flow, barometric pressure and humidity measurements are included in the enhanced Tester (Models 2H and 2G). The Tester is menu-driven and simple to operate. All functions are set from a 128 x 32 pixel graphical display. *not available in Europe.
Specifications
- General Information
- Temperature Requirements: 15° C to 35° C when operating 0° C to 50° C in storage
- Display/Control: LCD graphic display, 128 x 32 pixel
- Data Input/Outputs (1): Bi-directional RS-232C for Computer control
- Power: 9 Volt Alkaline Battery RG9 or Battery eliminator
- Case: ABS plastic case
- Weight: 0.4 kg / 0.9 lbs. with battery
- Dimensions (D x W x H): 34 mm x 94 mm x 156 mm (1.3 in x 3.7 in x 6.1 in)
- DPM4 Model 1H or 2H
- Operating Range : -350 mmHg to 350 mmHg
- Accuracy : 0.3 % of range
- Resolution : 0.1 mmHg
- Units of measure : mmHg, mBar, cmH 2 O, PSI, InHg, InH 2 O, kgcm 2 and kPa
- DPM4 Model 1G or 2G
- Operating range: : -700 mmHg to 5000 mmHg
- Accuracy : 0.3 % of range for temperatures from 21 º C to 25 ° C and relative humidity from 30 % to 70 %;, 0.3 % of range plus 0.02 % of range per degree C for temperature < 21 ° C or > 25 ° C with relative humidity from 30 % to 70 %
- Resolution: : 0.5 mmHg
- Units of measure: : mmHg, mBar, cmH 2 O, PSI, InHg, InH 2 O, kgcm 2 and kPa
- Temperature Measurement
- All Models (DPM4-1H, 1G, 2H, 2G)
- Operating range : -40 º C to 200 º C
- Accuracy : (2 % of reading, 0.5 º C)
- Resolution : 0.1 ° C and ° F
- Units of measure : ° C and ° F
- Use standard external temperature probe type PT-100 or PT-1000 (DIN/IEC 751 Class A) for temperature measurements in ° C or ° F
- PT-100 : -50 º C to +200 º C
- Accuracy : 0.23 º F @ 212 º F (0.13 º C @ 100 º C) 0.18 º F @ 32 º F (0.1 º C @ 0 º C) 0.36 º F @ 212 º F (0.2 º C @ 100 º C)
- PT-1000 : -50 º C to +200 º C 0.3 º C (0.5 º F)
- DPM4 Model 2G or 2H
- It is possible to compensate for the sea level and calibrate for offsets.
- Operating Range : 380 mmHg to 825 mmHg
- Accuracy : 2 % of reading 1LSD
- Resolution : 0.1 mmHg
- Units of Measure : mmHg, mBar, inHg and hPa
- Gas Flow : Measures with an embedded sensor with 11 calibration points to compensate nonlinearity. Calibration constants stored in firmware.
- Operating range : -750 to 750 ml/min
- Accuracy : 1% of range or 5% of reading, whichever is greatest
- Compatibility : 0.1 ml/min
- Resolution : Gas: Air, N 2 , O 2 , CO, NO, CO 2 , N 2 O, and NO 2
- Units of measure : ml/min (or SCCM - Standard Cubic Centimeters per Minute)
- Relative Humidity
- An integrated sensor in the instrument determines relative humidity measurements.
- Operating range 12 % to 95 % RH
- Accuracy 3.5 % of reading 2 % RH @ 25 º C for RH between 12 % and 95 % RH
- Resolution 0.1% RH
- Gas compatibility Air
- Units of measure % RH
Customer reviews