ISE -Model MIURA 200 -Fully-Automated Clinical Chemistry Analyser - Single / Double Arm

SHARE

The New Miura 200 Is An Auto-Ananlyser For The Clinical Chemistry, Immuno-Turbidimetry, Tdm And Doa Available Also With Special Application For Wine And Water Analysis.

Most popular related searches
  • The New Miura 200 is the idea we have for a compact bench top random access fully automated analyser with the performance of higher class instruments.
  • No compromise between throughput and size of the New Miura 200.
  • The New Miura 200 is the solution for all kinds of laboratories as a main unit or a back up unit.
  • 30/43 reagents on board plus diluent.
  • 49 samples on board.
  • Integrated ISE module for Na+, K+, Li+, Cl- in serum, plasma and diluted urine (optional).
  • Integrated cuvette washing station.
  • Integrated liquid level detector and probe shock sensor for samples and reagents.
  • Integrated positive identification for samples and reagents by barcodes.
  • Integrated micro metering diluter maintenance free.
  • 2 years warranty.
  • Miura 200 is 100% developed and produced in Italy.

  • Miura 200 Instrument Code :
  • MIURA 2SA (for single arm instrument)
  • MIURA 2DA (for double arm instrument)
  • System : Random access, batch, STAT
  • Throughput :
  • Double Arm
  • Up to 240 photometric tests per hour
  • Up to 420 tests hour with integrated ISE module (on request)
  • Single Arm
  • Up to 200 photometric tests per hour
  • Up to 360 tests hour with integrated ISE module (on request)
  • Robotic XYZ Arm : 1/ 2 arm equipped with liquid level detector (capacitive) and shock sensor
  • Analytical methods : End point mono or bichromatic, fixed time, kinetic, differential sample blank
  • ISE Module : Yes, Na+, K+, Li+, Cl- in serum, plasma and diluted urine (optional)
  • Number of reagent positions :
  • 30 parameters on line plus diluent with 50ml R1/ 20ml R2 vials /
  • 43 parameters on line plus diluent with 35ml R1/ 15ml R2 vials.
  • Reagents method : Mono, double
  • Number of samples / calibrators / controls : 49 sample/calibrator/control positions on primary tubes (Ø12÷13mm × 75/85mm) and sample cups (3ml)
  • Reagent volume : 1 ÷ 400 μl for Single Arm and 1 ÷ 300 μl for Double Arm
  • Sample volume : 1 ÷ 300 μl for Single Arm and 1 ÷ 500 μl for Double Arm
  • Typical reaction volume : 200/250 μl
  • Calibration curves : Linear/non linear curves up to 8 points, auto dilution of the calibration curve (linear regression, linear piece-wise, cubic spline, polynomial)
  • Reading cuvettes : 80 automatic re-washable
  • Cuvette washing : Automatic washing station
  • Diluter : High Precision Micro-metering pump to dispense reagents and samples (no syringe)
  • Automatic dilution : Automatic / on request
  • Reaction temperature : +37°C ± 0.1°C
  • Wavelengths : 8 from 340nm to 700nm plus one free position
  • Light source : Long life halogen lamp (2000 hours life)
  • Photometric linearity : 0 ÷ 3.0 OD
  • Water consumption : less than 3.5 liters / hr
  • Reagent cooler : Up to 14 °C below ambient temperature
  • Bar-code reader : Bar-code reader integrated for samples and reagents (optional)
  • Real-time electronics : Modular, SMD
  • Power supply : 100 ÷240 Vac
  • Quality control data : Control menu with Levy-Jennings plotted chart
  • Personal Computer : External, all in one (Touch screen PC)
  • Peripherals : CD-ROM, HD > 40 GByte
  • Printer : External laser
  • Operational system : MS Windows®
  • Interface to LIS host : Ethernet, Bi-directional connection
  • Environment temperature : 18 ÷ 32°C
  • Environment RH : 30 ÷ 70 % (without condensation)
  • Dimensions (WxDxH) : 93x74x60 cm
  • Weight : About 70 kg

Miura 200 software is designed to meet the requirements of a wide range of customers. Easy to use and flexible, it is a powerful solution for the customer who wants a quick solution for the daily laboratory management and for the customer who wants a high customizable system.

The flexible analyser enables easy addition of new methods. It is possible to program and store almost unlimited number of protocols, arranged on a panel drawn up by the user according to his needs. Execution is pure random access with the possibility to choose the batch mode.

MAIN SOFTWARE CHARACTERISTICS:

  • Programmable automatic stand-by and wake-up functionalities.
  • Intuitive and modern graphical user interface.
  • 3 levels of quality controls with Levy-Jennings graph.
  • Fully bidirectional L.I.S. support with or without Host/Query.
  • Automatic pre/post dilution of the samples.
  • Calibration with linear/nonlinear curves up to 8 points, auto dilution of the calibration curve (linear regression, linear piece-wise, cubic spline, polynomial).
  • Self-diagnosis and alarm procedures are an integral part of the software.
  • Powerful archive for patients, standard and Qc results.
  • Customizable profiles to easy manage the WL.