Bionime Rightest -Model GT100 -Glucose Meter
The GT100 makes testing easy and glucose results clear. Auto-Coding; Eliminates the need for manual coding and delivers maximum accuracy. Large Display : Easy to view results. Side-Insert Design : Easy to handle and prevents blood contact for excellent hygiene. Trusted Accuracy : Utilizes gold electrodes and a patented strip structure to provide accurate and reliable results. Tiny Blood Sample : Requires only 0.75µL.
Most popular related searches
blood glucose testing
blood glucose test
blood sampling
whole blood
blood sample capillary
glucose meter
blood glucose
blood testing
blood test
- Test strip ; GT100
- Coding ; Auto-coding
- Enzyme / Measurement technology ; GOD / Electrochemical sensor
- Blood sample source ; Capillary whole blood (finger, palm and forearm)
- Minimum sample volume ; 0.75 µL
- HCT range ; 30 - 57%
- Measuring range ; 0.6 - 33.3 mmol/L (10 - 600mg/dL)
- Meter dimensions (L x W x H) ; 98.0 x 46.0 x 17.5 mm
- Display area (L x W) ; 53.0 x 32.0 mm
- Weight (with battery) ; 53.0 ± 5 g
- Power supply ; 1 × CR2032 battery
- Memory capacity ; 500 blood glucose test results with date and time
- Operating temperature ; 10 - 40°C (50 - 104°F)
- Operating relative humidity ; 10 - 90%
- Meter storage conditions ; -10 - 60°C (14 - 140°F)
- Test strip storage conditions ; 4 - 30°C (39 - 86°F), < 90% relative humidity
- Accuracy ; 95% of the measured glucose values shall fall within either
- Glucose concentrations < 5.5 mmol/L (100 mg/dL), difference within ± 0.83 mmol/L (15 mg/dL)
- Glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), bias within ± 15%
- Precision ; Glucose concentrations < 5.5 mmol/L (100 mg/dL), SD < 0.28 mmol/L (5 mg/dL)
- Glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), CV < 5%