Esse3 -Model ES-HEM 300 -Haematology Analyzer

SHARE

Fully Auto Hematology Analyzer Is An Easy Operation . Blood Analytical System With 3-Part Differentiation Of Wbc. With Advanced Digital Technology, Float Discriminator, Intelligent Flush , Double Quantitative Technology To Ensure The Accuracy Of The Result.

Most popular related searches
  • 20 parameters and 3 color histograms ( WBC, RBC, PLT ) display
  • three detecting modes : VENOUS, CAPILLARY and PREDILUTES
  • Throughput 60 samples/H
  • Large 10.4’’ LCD colour display 640x480
  • Memory for storage upto 35.000 results ( including histograms) , L.J curver, X , SD , CV
  • Possibility to work in venous blood ( 9,6 μlt ) , capillary mode ( 9,6 μlt ) or prediluted ( 13 μlt )
  • Aperture diameter WBC/RBC : 100/80 μm
  • dilution rates : WBC whole blood1:304 ,capillary1:293 - RBC/PLT whole blood 1:44833 , capillary 1:44402
  • precision RBC + 2% - WBC + 2,5% - Hct et MGV + 2% - HGB + 1,5%
  • Carry over : WBC, RBC, HGB < 0.5% , PLT<1%
  • Calibration mode : Relatively independent manual calibration and auto-calibration in three detecting modes
  • Color Windows operating system, graphical buttons , supporting mouse operation
  • High precision microanalysis technology, low consumption of blood sample volume ( only 9,6 μlt )
  • Double convolution and intelligent fitting technology to ensure the accuracy of PLT counting
  • High voltage zap with monitor in real time, recoil flush technology
  • Automatically cleaning sample needle ( inside and outside)
  • Intelligent auto detection and maintenance functions
  • Intelligent float discriminator, manual adjustment
  • Automatic alarm for inadequate reagents and fully enclosed reagent tubing
  • Adjustable sample needles, suitable for various sample collection containers
  • Microprocessor controlled , Autotest and automatic calibration with automatic temperature compensation
  • Menu : Count, Sample Mode, Review, Quality Control, Setup, Service, Calibration, Help, Shutdown
  • Built-in thermal printer , paper wide 58 mm
  • multi-language SW , with 36 sets of alert messages
  • comprehensive QC programs including L-J and X-B analysis ; up to 9 control lots- 30 run/ lot
  • Outputs : 1 parallel interface, 2 PS/2 interfaces ( supporting networks) , 2 RS 232 ports , 2 USB , 1 VGA interface
  • Power supply 100 to 240V 50/60 Hz, consumption 220 VA
  • Environment : Temperature 15°C~35 °C – Humidity 30%~85%
  • Dimensions 55x50x60H cm, weight 34 kg