Yumizen - Model G1550 - Hemostasis Analyzer
The Yumizen G1550 automated analyzer meets those daily expectations thanks to its unique features: Clotting, immunologic and chromogenic assays, Integrated nephelometric and immuno-turbidimetric measuring channels, Up to 8 measuring channels with 2 wavenlenght combination, Continuous loading (sample racks, reagents, cuvettes), Real-time remote access, Comprehensive system (reagents, QC, consumables, services support), Safety single use cuvette, Touch screen, Full quality record for traceability (reagents lot, calibration, QC, rerun, users login...)
System auto-verification
- 44 cooled controlled positions for reagents
- Up to 4 buffer and 8 control positions
- Liquid level sensor
- Positive barcode identification
- On board reagents, consumables and liquid waste management
- Up to 1000 cuvettes for testing and 4.5L of cleaner
Run Calibration / QC
- Pre calibrated tests
- Auto-control setup, customizable according to activity (number of tests, new reagents...)
- Levey-Jennings quality control chart
Load Samples
- 150 samples / 15 racks
- Cap piercing (optional) for the Closed-tube mode
- 8 STAT sample positions
Request tests
- Worklist
- Random Access
- Comprehensive test pannel
Run and Validation
- Real time curve monitoring
- Automated rerun and reflex testing
- Measuring status feed back
Dimensions & Weight: Length: 1000 mm, Height: 600 mm, Depth: 900 mm; Weight: 85 Kg
External Printer: Available
Throughput: 240 Pt / hour
Sampling volume: 5 - 200µl
Temperature: 15°C - 32°C
Power: 100-240v 50Hz-60Hz, 5A
Power supply: external supply power
Export: 6 x USB plug
Screen: Color LCD touch screen 270x340 mm
Storage Capacity: 1.000.000 results
Parameter: Clotting tests (640 nm); Screening tests (PT, APTT, Fibrinogen, TT); FactorS (II, V, VII, X, VIII, IX, XI, XII); Immuno-turbidimetric test (570 nm); D-Dimer; Chromogenic test (405 nm); ATIII
Norms: EN 61326-2-6: 2006; IEC 61010-1: 2001; IEC 61010-2-101: 2002; EN ISO 13485: 2003 and 2012; NF EN ISO 14971: 2013
Directives: 98/79/EC (DIV); 2011/65/EU (RoHS); 1907/2006 (REACH); 2014/30/EU (Electro Magnetic Compatibility); 2014/35/EU (Voltage limits)