PaceTech
  1. Companies
  2. PaceTech
  3. Products
  4. Vitalmax - Model 4100SL - Medical ...

VitalmaxModel 4100SL - Medical Patient Monitors

SHARE

The VITALMAX 4100SL is designed with an optimal ability against electrosurgical interference. It offers a simple and adaptable interface. Patient information along with vital sign settings can be quickly modified to meet the needs of a patient’s changing condition. The VITALMAX 4100SL offers a high resolution 12″ color LCD touch screen for making quick screen adjustments, set default settings, alarm limits, and manage up to 72 hours of detailed patient data. It boasts a very short recovering time after defibrillation, especially suitable for use in an emergency or operating room.

Most popular related searches
  • 15-inch touch screen
  • Optional EtCO2, Dual IBP, Anesthetic Monitoring, 12-lead EKG
Plug in modules available with following configurations:
  • 2x IBP & CO
  • 2x IBP & EtCO2
  • EtCO2 Only
  • CO Only
Our line of patient monitors comes with the following standard options:
  • Masimo SpO2
  • 5-lead EKG (with additional 3-lead operation)
  • SunTech Non-Invasive Blood Pressure (NIBP)
  • Temperature Monitoring
  • Integrated Thermal Printer
Plug in modules available with following configurations:
  • 2x IBP & CO
  • 2x IBP & EtCO2
  • EtCO2 Only
  • CO Only
Our line of patient monitors comes with the following standard options:
  • Masimo SpO2
  • 5-lead EKG (with additional 3-lead operation)
  • SunTech Non-Invasive Blood Pressure (NIBP)
  • Temperature Monitoring
  • Integrated Thermal Printer

Physical Weight & Dimensions

  • Weight: 5kg
  • Height: ~11 inches
  • Depth: ~6-1/5 inches
  • Width: ~12 3/5 inches

Application

  • Neonatal, pediatric and adult patients

Operation Environment

  • Power Source - External AC power or internal battery
  • AC Power - 100-240VAC, 50/60Hz, <150VA
  • Battery - Rechargeable Lead-Acid; Operating time under the normal condition (1 hour); Operating time after the first alarm of low battery (10 min.)

Temperature

  • Working - 5°-40° C
  • Storage - -20°-65° C

Performance Specifications

  • Display - 12.1" color TFT, Touch Screen
  • Resolution - 800 × 600 pixels
  • Trace - 4, 6 or 8 waveforms ECG (I, II, III, aVR, aVL, aVF, V1-V6), PLETH, RESP, IBPx2, ETCO2
  • Indicator - Alarm indicator; Power indicator QRS beep and alarm sound
  • Trend Time - 72 hours
  • Recorder - Built-in, thermal array, 2 channels
  • Record - Width 48mm
  • Recorder - Paper 50mm
  • Record - Speed 25mm/s, 50mm/s

ECG

  • Input - 3-lead ECG cable and standard AAMI line for connection
  • Input - (Optional) 5-lead ECG cable and standard AAMI line for connection
  • Lead Choice - I, II, III, aVR, aVF, aVL, V
  • Gain Choice - x0.5, x1.0, x2.0
  • Frequency Characteristic - 0.05-40 Hz
  • ECG Waveforms - 7 channels
  • Penetration Voltage - 4000VAC 50/60Hz
  • Sweep Speed - 12.5, 25 and 50 mm/s
  • HR Display Range - 30-300bpm
  • Accuracy - ±1% or ±1bpm, whichever is greater
  • Alarm Limit Range Setting - Upper limit 80-400bpm; Lower limit 20-150bpm

RESP

  • Measure Method - RA-LL impedance
  • Range - 0-120 rpm
  • Accuracy - + 3 rpm
  • Alarm Upper-lower Limit Setting - Upper limit 6-120 rpm, Lower limit 3-120 rpm
  • Sweep Speed - 12.5 and 25 mm/s

NIBP (Suntech)

  • Measuring Technology - Automatic oscillating measurement
  • Cuff Inflating - <30s-0-300 mmHg, standard adult cuff
  • Measuring Period - AVE<40s
  • Mode - Manual, Auto, STAT
  • Measuring Interval in AUTO Mode - 2 min-4 hrs
  • Pulse Rate Range - 30-250 (bpm)
  • Measuring Range - Adult/Pediatric Mode, SYS 40-250 (mmHg), DIA 15-200 (mmHg); Neonatal Mode, SYS 40-135 (mmHg), DIA 15-100 (mmHg)
  • Resolution - 1mmHg
  • Accuracy - Pressure, Maximum Mean error ±5mmHg, Max. Standard deviation 8mmHg
  • Overpressure Protection - Adult Mode 280(mmHg), Neonatal Mode 150 (mmHg)
  • Alarm Limit Setting - SYS 50-240 mmHg, DIA 15-180 mmHg

TEMP

  • Range - 25°-50° C
  • Accuracy - + 0.2 °C 25.0°-34.9° C; + 0.1° C 35.0°-39.9° C; + 0.2° C 40.0°-44.9° C; 0.3° C 45.0°-50.0° C
  • Display Resolution - 0.1° C
  • Alarm Upper-lower Limit Setting - Upper limit 0-50° C, Lower limit 0-50° C
  • Channel - 2 channels
  • Alarm Limit Setting - 10-50° C

Masimo SET SpO2

  • SpO2 Accuracy (non-motion):
  • Adult Pediatric: 70~100 %: ±2 %, 0~69%: unspecified
  • Neonate: 70~100 %: ± 3 %,0~69 %: unspecified
  • SpO2 Accuracy (motion):
  • Adult Pediatric: 70~100 %: ±3 %, 0~69%: unspecified
  • Neonate: 70~100 %: ± 3 %, 0~69 % : unspecified
  • Low Perfusion Performance: >0.02 % Pulse Amplitude and % Transmission >5
  • SpO2: ±2 %
  • PR: ±3 bpm
  • Modes:
  • Averaging mode: 2,4,8,10,12,14 and 16 s
  • Sensitivity: Normal, APOD and Maximum
  • PR Accuracy (non-motion):
  • Neonate: 70~100 %: ± 3 %, 0~69 % : unspecified
  • Adult Pediatric Neonate: 25~240 bpm: ±3 bpm
  • PR Accuracy (motion):
  • Adult Pediatric Neonate: 25~240 bpm: ±5 bpm
  • Measuring Range:
  • SpO2: 1~100 %
  • PR: 25~240 bpm
  • Perfusion: 0.02~20 %

IBP (Option)

  • Measurement Range - -10-300mmHg
  • Channel - 2 channels
  • Pressure Transducer Sensitivity, 5μV/V/mmHg
  • Impedance Range - 300-3000Ω
  • Transducer Sites - ART, PA, CVP, RAP, LAP, ICP
  • Resolution - 1mmHg
  • Accuracy - ±1mmHg or ±2%, whichever is greater
  • Alarm range - -10-300mmHg

EtCO2 (Option)

  • EtCO2 Option allows compliance with
  • Mode of Sampling – Sidestream
  • Principle of Operation -Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts
  • CO2 Measurement Range - 0 to 150 mmHg (0 to 19.7%, 0 to 20 kPa)
  • CO2 Calculation Method - BTPS (Body Temperature Pressure Saturated)
  • CO2 Resolution - 0.1mmHg(0-69mmHg); 0.25mmHg(70-150mmHg)
  • CO2 Accuracy - 0-40 mmHg ± 2 mmHg; 41-70 mmHg ± 5% of reading; 71-100 mmHg ± 8% of reading; 101-150 mmHg ± 10% of reading; Above 80 breath per minute ± 12% of reading
  • Sampling Rate - 100Hz
  • Respiration Rate - 2-150 bpm
  • Respiration Rate Accuracy - ±1 breath
  • Response Time - <3 seconds - includes transport time and rise time
  • Inspired CO2 Measurement Range - 3-50 mmHg