MEDICOR BabyLife - Model BLR-2100A - Warming and Resuscitation Table
Automatic warm-up. Pre-warm (33 oC) & Manual mode (0-100% in 5% increments). Surface temperature servo control (30-38 oC). Skin temp. servo control (32-38 oC increments 0.1 oC) with reusable sensor. Heavy duty stand with 4 castors (2 lockable, 150 mm). Protector bumpers on each side & handle on front. Foldable side walls (4 pcs, height 17,5 cm) w/ name plate. Tube ports & measure line on side walls.
-
Most popular related searches
Featured Benefits
- Lying surface w/ antidecubitus mattress, x-ray compatible
- Cradle tilting (continuous, up to 10 degrees, both direction)
- Drawable x-ray cassette tray (35 x 65 cm)
- Turnable head by 30 and 90 degrees both direction
- Ceramic heaters with parabolic reflectors and protecting grid
- Cupboard with two drawers (300x300x90 mm)
- Gas cylinder holders (two) on the stand
- Monitor shelves (two) & IV pole on the column
- Dimmable halogen & LED procedure light
- 5” LCD display with APGAR timer
- Set / Display of date / time / temperature / trend, etc.
- Water resistant, foil-covered keypad with lock
- State-of-the-art microprocessor controller
- Visual and audible alarms and backup sensors
- Continuous self-test function during operation
- Test button / Self test / Silence / Reset button
- RS-232 connector
- 230V/50Hz+/-10% (230V/60Hz or 110V/50-60Hz on request)
Technical Data
- PHYSICAL ATTRIBUTES (STANDARD VERSION)
- Height / Adjustable / Packing : 185/185-200/112 cm (72,8/72-78,7/44,1 in)
- Width / Packing : 120/127 cm (47,2/50 in)
- Depth / Packing : 71/81 cm (28/31,9 in)
- Bed to Floor (fix) : 87 cm (33,5 in)
- Bed to Floor (adjustable) : 87 – 102 cm (33,5 – 39,4 in)
- Weight / Packing approx. : 90/120 kg (199/265 lbs)
- Mains voltage : 220/230V ±10% (default), 110/120V ±10% (upon request)
- Mains frequency : 50 Hz (default), 60 Hz (upon request)
- Nominal power consumption : 1100W
- Nominal power of ceramic heaters (2 pcs) : 500 W
- Operating ambient temperature range : +20°C - +30°C (68oF – 86oF)
- Storage/transport ambient temperature range : -40oC - +70oC (-40oF - 158oF)
- Operating (Storage) ambient RH range : 30% - 75% (10% - 100%)
- Operating (Storage) air pressure range : 700 hPa - 1060 hPa (500 hPa - 1060 hPa)
- LYING SURFACE (BED) SPECIFICATION
- Surface size : 61 x 76 cm (24 x 29,9 in)
- Mattress size : 60 x 75 x 2 cm (23,6 x 29,5 x 0,8 in)
- Mattress to head height : 80 cm (31,5 in)
- TEMPERATURE CONTROL MODES
- Pre-warm mode : 33.0°C (91.4°F)
- Manual control temperature mode : 0%-100%, in 5% increments
- Surface control temperature range : 30.0°C (86°F) to 38.0°C (100.4°F), in 0.1°C increments
- Skin control temperature range : 32.0°C (89.6°F) to 38.0°C (100.4°F), in 0.1°C increments
- Mattress heating temperature range : 30.0°C (86°F) to 33.0°C (91.4°F), in 1°C increments
- Peripherial skin / Rectal temperature monitoring : Yes
- PERFORMANCE
- Temperature rise time : 25 °C (77°F) ambient < 30 min
- Temperature variability : under 0.5°C (0.9°F)
- Temperature uniformity on the surface of the mattress : under 0.7°C (1.2°F)
- Correlation of the indicated air temperature with the actual surface temperature : ≤ 0.7°C (1.2°F)
- Display precision of skin temperature : under ±0.1°C (0.1°F)
- Measuring precision of skin temperature sensor : under ±0.3°C (0.5°F)
- APGAR warning signal after : 1 / 5 / 10 min
- OXYGEN SERVO CONTROL
- Oxygen control range : 21% to 90% (head hood)
- Oxygen control accuracy of full scale : ± 3%
- Oxygen display accuracy (100% calibration) : ± 3%
- Oxygen display accuracy (21% calibration) : ± 3%
- Oxygen display resolution : 1%
- WEIGHING SCALE
- Weight range : 0 to 8 kg (17,1 lbs)
- Weight display resolution : 5 gr or 0,2 oz
- Weight accuracy up to : 5 kg ± 5 gr, over 5 kg ± 10 gr
- Low voltage alarm :
- LOADING
- Max. load of mattress : 10 kg (22,1 lbs)
- Max. load of IV pole : 2 kg (4,7 lbs)
- Max load of shelves : 2 kg (4,7 lbs)
- DEVICE CLASSIFICATION
- Protection class : Class I, Type BF, continuous operation
- Ingress of liquids : IP20
- Electric safety : IEC 60601-1; IEC 60601-2-21
- TREND PARAMETERS
- 3/24/170-hour trend : Surface temperature, Skin temperature, Mattress temperature, Peripherial temperature, Oxygen concentration, Heater power