MEDPLANT - Model Oxitest-1 - Portable Pulse Oximeter with Adult Sensor
From Pulse Oximeters
Field of application: emergency medical services, disaster medicine services, military and extreme medicine services; anesthesiology departments, ICUs and resuscitation departments; maternity hospitals, medical ambulance, etc. Current use:Pulse oximeter Oxitest-1 is used for non-invasive continuous monitoring saturation and pulse rate.
Technical characteristics
SpO2 saturation
Indicating range (%)from 0 to 99
Measurement limits (%)85-99
60-84
less than 60
Measurement inaccuracy (%)
from 85 to 992% (absolute accuracy)
from 60 to 843% (absolute accuracy)
less than 60isn’t normative
Indicating range (1/min.)20-250
Measurement limits (1/???.)30-120
120-250
Measurement inaccuracy (1/min.)
30-120± 1
120-250± 2
Indication
7-segment LED of saturation2 pcs, size 21?14 mm
7-segment LED of pulse3 pcs, size 14?10 mm
Linear scaleIndication of pulse wave, perfusion level
Reaction time for parameters changes10 sec
Four-level indication of the accumulator chargingComplete charge
half charge
weak charge
almost full discharge
Changeable brightness levelfrom 21 to 0 mcd
Alarm
Fixed saturation and pulse level
Signal quality
Sensor absence
Absence of finger in sensor (bad signal)
Range of limits, SpO2
- lower limitfrom 50 to 99 with pitch (± 1)
- upper limitfrom 80 to 99 with pitch (± 1)
Range of limits, PR
- lower limit (min/1)from 30 to 200 with pitch 5
- upper limit (min/1)from 70 to 250 with pitch 5
Separate volume controlUpon “alarm” or pulse
Power supply
Combined power
From three batteries?? Type
From three batteries of disposable usage?? Type
From on-board's power system, V8-18
From main supply, V; Hz220; 50
Self-contained operation from the battery50 hours
Other
Overall dimensions, mm137?64?27 ±10
Weight, kg0,3
Advantages
- Electronic security from power surges
- Electronic security from incomplete battery charge
- Stable saturation indication at low pulse strength and move actions
- Alarm system for signal quality, probe absence, finger absence in the probe
- Separate volume control of alarm signal and pulse
- Three typed of power supply: ???, accumulator, mains supply
- Two probes are available
- Steel cramp for fixation on the car console
- Large and bright indication
Customer reviews
No reviews were found for MEDPLANT - Model Oxitest-1 - Portable Pulse Oximeter with Adult Sensor. Be the first to review!