1. products
  2. random access clinical chemistry autoanalyzer

Diconex -Model InCCA Bit -Random Access Clinical Chemistry Autoanalyzer

SHARE

InCCA Bit is a random access clinical chemistry autoanalyzer for small volume laboratories; this analyzer will ensure the requirements of an affordable automated system. It is an open system, utilizes PMMA washable and reusable cuvettes, with friendly software interface. Lowest water consumption, ideal to start your laboratory automatization. Low maintenance units, having cycle counters for parts, 100% cost-effective.

Most popular related searches

GENERAL

  • Average throughput: 100 tests/hour – 300 test/hour with ISE
  • Random access, automated, discrete
  • Open system

SOFTWARE

  • Stats, samples, calibrators and controls loading at any time in the process
  • Unlimited number of methods in memory
  • Calibration curves and reaction curves
  • Automatic re-dilution and repetition
  • Reaction graph
  • Preventive maintenance alarms
  • Levy – Jennings plots
  • LIS
  • Reagent inventory check
  • QC graphics
  • Settings for DOA and TDM test
  • Linux and Windows compatible
  • Analyzer testing
  • Reagent statistics consumption

OPTICAL SYSTEM

  • Double beam
  • Interferential filters: 340, 380, 405, 450, 505, 546, 578, 600, 650, 700 nm
  • Long life halogen lamp
  • Optical path: 6 mm
  • Photometric range: -0.100 to 3.600 A
  • Measurement type: monochromatic and bichromatic
  • Direct reading for several solutions

PROBE HYDRAULIC SYSTEM

  • High resolution dilutor
  • Minimun sample volume: 2 µl
  • Minimun reading volume: 220 µl

ANALYSIS MODES

  • End point reagent blank
  • End point sample blank
  • Kinetics
  • Fixed time kinetics
  • Cut off

METHODS

  • Colorimetric
  • Turbidimetric
  • Kinetics
  • Ion selective electrode

POWER REQUIREMENTS

  • 100/240 VAC 50/60 Hz.

MEASURES

  • Width: 350 mm Height: 470 mm Depth: 380 mm.
  • Weight - 20 kg.

SAMPLE AND REAGENT PIPETING SYSTEM

  • Preheated probe with additional mixer integrated
  • Internal and external probe washing
  • Conductive probe level sensor
  • Impact sensor (vertical & horizontal)

SAMPLE AND REAGENT TRAY

  • 22 sample position
  • 24 reagent position
  • Primary tubes 12/13 x 75/100 mm, pediatric cups
  • Continuous sample loading

CUVETTE TRAY

  • 100 reusable PMMA extra longlife reaction cuvettes
  • Warm air bath
  • Working temperature: 37º C (Variable from room T +/- 5ºC to 45ºC)

OPTIONAL

  • Sample bar - code reader
  • Extra filter: any wavelength
  • ISE module (Na+, K+, Cl-, Li+)