Airon Corporation
  1. Companies
  2. Airon Corporation
  3. Products
  4. pNeuton - Model S - Ventilator

pNeutonModel S - Ventilator

SHARE

pNeuton model S Ventilator with CPAP for EMS Ground and Air Transport. pNeuton model S ventilator provides mask CPAP plus volume or pressure ventilation for pediatric to adult patients for pre-hospital and hospital environments; with simple-to-use controls – Tidal Volume, Respiratory Rate and Pressure Limit. For early intervention, the built-in CPAP (Continuous Positive Airway Pressure) system allows you to start with face mask CPAP and if the patient needs it, seamlessly provide ventilation without interruption. Fully pneumatic, the pNeuton S does not use batteries or electrical power to provide ventilation with readiness for operation.

Most popular related searches

Meeting the Needs for the Most Challenging Patients in Respiratory Distress
The built-in high output demand flow CPAP system with automatic sensitivity insures critical care ventilation needs are optimized for spontaneous and mandatory breaths.

Benefits for the Patient
  • Provides invasive or non-invasive patient support
  • Low patient Work of Breathing with peak flow of up to 140 L/min to meet patient needs
  • 100% or 65% to match patient oxygen requirements
  • Switch from mask CPAP to invasive ventilation seamlessly and without interruption
  • Low-profile, easy-seal mask reduces facial pressure
Benefits for the Healthcare Provider
  • Calibrated CPAP and oxygen controls for easy operation
  • MRI conditional up to 3 T – place right next to the scanner with remote alarm output for the control room
  • Manometer to monitor patient pressure and effort
  • Built-in CPAP / PEEP system with automatic sensitivity for changing patient needs; no need for external CPAP valves
  • Night visible pressure gauge for patient monitoring
  • Monitor CO2 with invasive/non-invasive ventilation
  • Oxygen powered – no batteries to change
  • Rugged and lightweight (6 lbs / 2.7 kg)
  • Use the bacteria / viral filter to protect caregiver from patient exhalation
Benefits for Your Budget
  • Two oxygen levels; conserve oxygen with 65% setting
  • 45-minute CPAP run time on a D cylinder
  • Demand flow CPAP system with minimal wasted oxygen
  • Economical patient circuits save on operating costs
  • No batteries or electronics decrease maintenance cost

pNeuton Model S P/N 20002
pNeuton Model S StartUp Kit P/N 20012

Description
  • Pneumatically operated for use on pediatric and adult patients > 23 kg
  • Demand flow system for spontaneous breathing to 140 L/min
  • Pressures displayed on manometer
  • Modes of Operation: CMV, IMV, CPAP, Pressure Limit
  • Mandatory breath flow rate is fixed at 36 L/min
  • MRI conditional – 3 T or less, maximum spacial gradient magnetic field of 720-gauss/cm or less, no gauss line restriction
  • Weight: 6 pounds (2.7 kg)
  • Dimensions: 4″ H x 8″ W x 6″ D (10 x 20 x 15 cm)
  • Input gas requirement: 55 psi ± 15 psi (3.8 bar ± 1 bar)
  • Classification according to MDD 93/42/EEC: IIb
  • Meets International Standards for transport ventilators:
    • ASTM: Ventilators Intended for Use in Critical Care (F 1100-90)
    • ISO: Lung Ventilators for Medical Use, Part 3 (EN 794-3:1999)
Control Settings
  • Mandatory Breaths: ON or OFF
  • Respiratory Rate: 2 to 50 bpm
  • Tidal Volume: 360 to 1,500 ml
  • Peak Pressure: 10 to 75 cm H2O (mbar)
  • PEEP / CPAP: 0 to 20 cm H2O (mbar)
  • Oxygen: 100% or 65%
Audible and Visual Alarms
  • All pneumatic alarm system (no batteries)
  • Low gas source pressure
  • If source gas drops below 30 psi (2 bar)
  • Continues operation as long as oxygen is available