Elemental Microanalysis Ltd.
- Home
- Companies
- Elemental Microanalysis Ltd.
- Videos
- How to Pack a CHN Oxidation and ...
How to Pack a CHN Oxidation and Reduction Reactor Video
Packing a single tube CHN oxidation and reduction reactor. Before adding the reagents it is helpful to mark up the heights of the individual zones using a marker pen, then clamp the tube vertically. Put in the first layer of quartz wool, or quartz discs, packing this down firmly with a suitable rod to make sure no particles may subsequently slip through.